Measurements
R&S
®
MXO 4 Series
214User Manual 1335.5337.02 ─ 05
Meas. type Symbol Description/result
Cycle max CMax Maximum value of one cycle
Cycle min CMin Minimum value of one cycle
Cycle peak to
peak
CPTP Peak-to-peak value of one cycle: the difference of CMax and CMin
9.2.4.4 Counting
Meas. type Symbol Description/result
Pulse count PCount The number of positive or negative pulses of the waveform, or of both positive and negative pul-
ses.
The mean value of the signal is determined. If the signal passes the mean value, an edge is coun-
ted. A positive pulse is counted if a rising edge and a following falling edge are detected. A nega-
tive pulse is counted if a falling edge and a following rising edge are detected.
Edge count EdCo The number of positive or negative edges, or of both positive and negative edges.
The instrument determines the mean value of the signal and counts an edge every time the signal
passes the mean value.
9.2.5 Settings for measurements
In the "Measurement" dialog, you can enable measurement and define their settings.
9.2.5.1 Setup tab
Access: "Menu" > "Measurement" > "Setup" tab.