Remote control commands
R&S
®
MXO 4 Series
677User Manual 1335.5337.02 ─ 05
Usage: Asynchronous command
Manual operation: See "Data setup: Condition, Pattern, Position" on page 341
TRIGger:LIN:ICONdition <IdOperator>
Sets the operator to set a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one identifier
pattern to be set with TRIGger:LIN:IMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger:LIN:IMIN and TRIGger:LIN:IMAX.
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Identifier setup: Condition, Frame ID min, Frame ID max"
on page 341
TRIGger:LIN:IMAX <IdPattern>
Sets the end value of an identifier range if TRIGger:LIN:ICONdition is set to
INRange or OORange.
Parameters:
<IdPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Identifier setup: Condition, Frame ID min, Frame ID max"
on page 341
TRIGger:LIN:IMIN <IdPattern>
Specifies a secondary identifier pattern, or sets the start value of an identifier range.
Parameters:
<IdPattern> List of comma separated values
Usage: Asynchronous command
Manual operation: See "Identifier setup: Condition, Frame ID min, Frame ID max"
on page 341