Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 460

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
460User Manual 1335.5337.02 ─ 05
Suffix:
<se>
.
1..3
2
Parameters:
<Events> Range: 1 to 2147483647
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "B event count" on page 137
TRIGger:MEVents:SEQuence<se>:DELay
<Delay>
Sets the time that the instrument waits after an A-trigger until it recognizes B-triggers.
Suffix:
<se>
.
1
Parameters:
<Delay> Range: 0 to 50
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Delay" on page 137
TRIGger:MEVents:SEQuence<se>:RESet:EVENt <State>
If enabled, the trigger sequence is restarted by the R-trigger condition if the specified
number of B-triggers does not occur before the R-trigger conditions are fulfilled.
Suffix:
<se>
.
3
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Enable reset event" on page 138
TRIGger:MEVents:SEQuence<se>:RESet:TIMeout:TIME <ResetTimeout>
The time the instrument waits for the number of B-events specified using TRIGger:
MEVents:SEQuence<se>:COUNt, before the sequence is restarted with the A-trig-
ger.
Suffix:
<se>
.
Irrelevant, omit the suffix.
Trigger