Remote control commands
R&S
®
MXO 4 Series
710User Manual 1335.5337.02 ─ 05
Parameters:
<Frequency> Range: 0.001 to 1000000
Increment: 1
*RST: 1000
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Frequency" on page 366
WGENerator<wg>:MODulation:FM:SYMMetry <RampSymmetry>
Sets the symmetry for the FM ramp modulation waveform, the percentage of time that
the waveform is rising.
Suffix:
<wg>
.
1..2
Parameters:
<RampSymmetry> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Symmetry" on page 366
WGENerator<wg>:MODulation:FM[:FUNCtion] <SignalType>
Selects the type of the modulating signal for FM modulation.
Suffix:
<wg>
.
1..2
Parameters:
<SignalType> SINusoid | SQUare | RAMP
*RST: SINusoid
Usage: Asynchronous command
Manual operation: See "Signal type" on page 365
WGENerator<wg>:MODulation:FSK:FONE <Frequency1>
WGENerator<wg>:MODulation:FSK:FTWO <Frequency2>
Sets the frequency of the first /second signal in FSK modulated signal.
Suffix:
<wg>
.
1..2
Waveform generator (option R&S MXO4-B6)