Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 376

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Waveform generator (option R&S MXO4-B6)
R&S
®
MXO 4 Series
376User Manual 1335.5337.02 ─ 05
Generating an arbitrary waveform from a saved file
If you want to output single periods of the arbitrary waveform, first set up the generator,
and then connect the DUT.
1. Tap the "Gen" signal activator.
2. Select the tab of the generator that you want to configure: "Gen1" or "Gen2".
3. In the "Setup" tab, set "Function type" = "Arbitrary".
4. Select "Arbitrary Setup" to open the arbitrary dialog.
5. Set the "Arb wfm source" to "Saved waveform".
6. Set the "Sample rate" if it is not defined in the file data.
7. Check "Number of samples".
8. Select "Back" to return to the "Setup" tab.
9. Select the "Run mode".
10. If "Run mode" = "Repetitive", set "State" = "On" to output the waveform at the out-
put connector of the waveform generator.
11. If "Run mode" = "Single period":
a) Connect the DUT.
b) Set "State" = "On" to enable the output of the waveform generator.
c) Tap "Run single" to output one waveform.
Generating an arbitrary waveform from the scope waveform
1. Tap the "Gen" signal activator.
2. Select the tab of the generator that you want to configure: "Gen1" or "Gen2".
3. In the "Setup" tab, set "Function type" = "Arbitrary".
4. Select "Arbitrary Setup" to open the arbitrary dialog.
5. Set the "Arb wfm source" to "Scope waveform".
6. Select the "Signal source", the waveform source for the generated waveform.
7. Tap "Update" to load the selected waveform.
8. Press "Back" to return to the "Setup" tab.
9. Select the "Run mode".
10. Set "State" = "On" to enable the output of the generated waveform.
11. If "Run mode" = "Single period", tap "Run single" to output one waveform.
Configuring the waveform generator
Loading ...
Loading ...
Loading ...