Remote control commands
R&S
®
MXO 4 Series
532User Manual 1335.5337.02 ─ 05
MEASurement<mg>:AMPTime:PTCount <PulseCount>
Sets the number of positive pulses for the pulse train measurement. It measures the
duration of N positive pulses from the rising edge of the first pulse to the falling edge of
the N-th pulse.
Suffix:
<mg>
.
1...16, index of the measurement
Parameters:
<PulseCount> Range: 1 to 2147483647
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Pulse count" on page 216
MEASurement<mg>:AMPTime:PSLope <PulsesSlope>
Sets the first slope of the pulses to be counted. The setting is only relevant for pulse
count measurement (MEASurement<mg>:MAIN is set to PULCnt.
Suffix:
<mg>
.
1...16, index of the measurement
Parameters:
<PulsesSlope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Pulse slope" on page 216
MEASurement<mg>:AMPTime:DELay<n>:DIRection <EdgeCountDir>
Selects the direction for counting slopes for each source: from the beginning of the
waveform, or from the end.
Suffix:
<mg>
.
1...16, index of the measurement
<n> 1..2
Selects the source number.
Parameters:
<EdgeCountDir> FRFI | FRLA
FRFI - FRom FIrst, counting starts with the first edge of the
waveform.
FRLA - FRom LAst, counting starts with the last edge of the
waveform.
*RST: FRFI
Usage: Asynchronous command