Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 535

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
535User Manual 1335.5337.02 ─ 05
Return values:
<Stop> Range: -1E+26 to 1E+26
Increment: 0
*RST: 0
Usage: Query only
Asynchronous command
MEASurement<mg>:RESult:EVENts:COUNt?
Returns the number of measured events in one acquisition.
The command is relevant for measurements of all events, see MEASurement<mg>:
MULTiple.
Suffix:
<mg>
.
1...16, index of the measurement
Return values:
<Count> Number of events
Usage: Query only
Asynchronous command
MEASurement<mg>:RESult:EVENts:STARt? <EventIndex>
Returns the start time of the indicated measured event.
The command is relevant for measurements of all events, see MEASurement<mg>:
MULTiple.
Suffix:
<mg>
.
1...16, index of the measurement
Setting parameters:
<EventIndex> Index number of the measured event.
Return values:
<EventStart> Range: -1E+26 to 1E+26
Increment: 0
*RST: 0
Usage: Query only
Asynchronous command
MEASurement<mg>:RESult:EVENts:STOP? <EventIndex>
Returns the end time of the indicated measured event.
The command is relevant for measurements of all events, see MEASurement<mg>:
MULTiple.
Suffix:
<mg>
.
1...16, index of the measurement
Automatic measurements
Loading ...
Loading ...
Loading ...