Remote control commands
R&S
®
MXO 4 Series
500User Manual 1335.5337.02 ─ 05
Manual operation: See "Operator" on page 175
CALCulate:MATH<m>:VERTical:SCALe[:VALue] <VerticalScale>
Sets the scale of the y-axis in the math function diagram. The value is defined as
"<unit> per division", e.g. 50
mV/div. In this case, the horizontal grid lines are displayed
in intervals of 50 mV.
Suffix:
<m>
.
1...5, index of the math waveform
Parameters:
<VerticalScale> Range: 1E-12 to 100E+12
Increment: 1E-05
*RST: 0.5
Default unit: V
Usage: Asynchronous command
Manual operation: See "[Scale]" on page 39
CALCulate:MATH<m>:VERTical:OFFSet <VerticalOffset>
Sets a voltage offset to adjust the vertical position of the math function on the screen.
Negative values move the waveform up, positive values move it down.
Suffix:
<m>
.
1...5, index of the math waveform
Parameters:
<VerticalOffset> Range: -100E12 to 100E12
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
Manual operation: See "Vertical offset" on page 178
CALCulate:MATH<m>:ENVSelection <EnvelopeCurve>
Selects the upper or lower part of the input waveform for mathematic calculation, or a
combination of both.
Suffix:
<m>
.
1...5, index of the math waveform
Parameters:
<EnvelopeCurve> MIN | MAX | BOTH
*RST: BOTH
Usage: Asynchronous command
Manual operation: See "Envelope wfm selection" on page 176