Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 577

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
577User Manual 1335.5337.02 ─ 05
FRANalysis:MEASurement:RBW <RBW>
Sets the resolution bandwidth, which determines the number of measurements that are
used for creating the plot.
Parameters:
<RBW> HIGH | MID | LOW
*RST: MID
Usage: Asynchronous command
Manual operation: See "RBW" on page 248
FRANalysis:OUTPut[:SOURce] <OutputChannel>
Sets the channel for the output signal of the DUT.
Parameters:
<OutputChannel> C1 | C2 | C3 | C4
*RST: C2
Usage: Asynchronous command
Manual operation: See "Output" on page 244
FRANalysis:POINts:LOGarithmic <PtsPerDecade>
Selects the number of points that are measured per decade, if FRANalysis:POINts:
MODE is set to DECade.
Parameters:
<PtsPerDecade> Range: 10 to 500
Increment: 1
*RST: 10
Usage: Asynchronous command
Manual operation: See "Points" on page 245
FRANalysis:POINts:MODE <PointsMode>
Selects, if the number of points for the FRA are measured as total or per decade.
You can set the number of points with FRANalysis:POINts:TOTal/ FRANalysis:
POINts:LOGarithmic.
Parameters:
<PointsMode> TOTal | DECade
*RST: DECade
Usage: Asynchronous command
Manual operation: See "Points" on page 245
Applications
Loading ...
Loading ...
Loading ...