Rohde & Schwarz MXO44-245 Oscilloscope

User Manual - Page 721

For MXO44-245.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-245 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
721User Manual 1335.5337.02 ─ 05
Bits:
0 = ALIGnment
2 = AUToset
3= WTRIgger (wait for trigger)
4= MEASuring
Usage: Query only
STATus:OPERation:ENABle <Enable>
Controls the ENABle part of the STATus:OPERation register. The ENABle defines
which events in the EVENt part of the status register are forwarded to the OPERation
summary bit (bit 7) of the status byte. The status byte can be used to create a service
request.
Parameters:
<Enable> Range: 1 to 65535
Increment: 1
Example:
STATus:OPERation:ENABle 5
The ALIGnment event (bit 0) and AUToset event (bit 2) are for-
warded to the OPERation summary bit of the status byte.
17.19.3 STATus:QUEStionable registers
The commands of the STATus:QUEStionable subsystem control the status reporting
structures of the STATus:QUEStionable registers.
See also: Chapter 16.5.3.4, "STATus:QUEStionable register", on page 393.
The query of the [:EVENt] and [:ENABLe] commands, return a list of the affected
sources:
Table 17-8: Source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:
[:ENABLe]
Bits Source values Description
For all bits
NONE
ALL
All available sources are activated
PPSupply PRobe<m>
Active probe
POVerload PRobe<m>
Active probe
COVerload CHAN<m>
Analog channels
WARNCHAN<m>
Warning overload channels
EXTTRIGGERIN
External analog signal connected to the external
trigger input
TRIGGEROUT
Trigger out signal
Status reporting
Loading ...
Loading ...
Loading ...