Remote control commands
R&S
®
MXO 4 Series
404User Manual 1335.5337.02 ─ 05
*CAL?...........................................................................................................................404
*CLS.............................................................................................................................404
*IDN?............................................................................................................................404
*OPC............................................................................................................................405
*OPT?...........................................................................................................................405
*RCL.............................................................................................................................405
*RST.............................................................................................................................405
*SAV.............................................................................................................................406
*SRE............................................................................................................................ 406
*STB?...........................................................................................................................406
*TRG............................................................................................................................ 406
*WAI............................................................................................................................. 406
*CAL?
Starts a self-alignment of the instrument, and then queries a status response. Return
values 0 indicate an error.
Return values:
<State> 0: no error
1: alignment failed
2: not aligned, e.g. init
3: device needs longer warmup time before selfalignment can
start
4: input signal connected during selfalignment
Usage: Query only
*CLS
Clear status
Sets the status byte (STB), the standard event register (ESR) and the EVENt part of
the QUEStionable and the OPERation registers to zero. The command does not
alter the mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
*IDN?
Identification
Returns the instrument identification.
Return values:
<ID> "Rohde&Schwarz,<device type>,<part number>/<serial num-
ber>,<firmware version>"
Example:
Rohde&Schwarz,MXO4,1335.5050K04/100222,1.00.0.2
Model assignment:
Usage: Query only
Common commands