Remote control commands
R&S
®
MXO 4 Series
603User Manual 1335.5337.02 ─ 05
MOSI
Triggers on a specified data pattern in that is expected on the
MOSI line.
MISO
Triggers on a specified data pattern in that is expected on the
MISO line.
*RST: FRSTart
Usage: Asynchronous command
Manual operation: See "Type" on page 290
TRIGger:SPI:DMINpattern
<DataPattern>
Specifies a data bit pattern, or sets the start value of a pattern range.
Parameters:
<DataPattern>
Usage: Asynchronous command
Manual operation: See "Data" on page 291
TRIGger:SPI:DPOSition <DataPosition>
Sets the number of bits or words to be ignored before the first bit or word of interest.
The effect is defined by TRIGger:SPI:PALignment.
Parameters:
<DataPosition> Range: 1 to 4096
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Position" on page 291
TRIGger:SPI:FCONdition <DataOperator>
Selects the operator for the MISO and MOSI pattern.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Data" on page 291