Trigger
R&S
®
RTB2000
67User Manual 1333.1611.02 ─ 03
time
Figure 5-8: Timeout trigger with range Stays
High
TRIGGER > "Trigger Type" = "Timeout"
Figure 5-9: Timeout trigger menu
Range
Selects the relation of the signal level to the threshold:
Stays High
The signal level stays above the trigger level.
Stays Low
The signal level stays below the trigger level.
Stays High|Low
The signal level stays above or below the trigger level.
Remote command:
TRIGger:A:TIMeout:RANGe on page 275
Time
Defines the time limit for the timeout at which the instrument triggers.
Remote command:
TRIGger:A:TIMeout:TIME on page 275
Threshold
Threshold of the trigger source channel, used as trigger level for the timeout trigger.
See also "Threshold" on page 44 and "Trigger Level, Threshold" on page 58.
Timeout Trigger
Loading ...
Loading ...