Rohde & Schwarz RTB2002 Oscilloscope

User Manual - Page 294

For RTB2002.

PDF File Manual, 510 pages, Read Online | Download pdf file

RTB2002 photo
Loading ...
Loading ...
Loading ...
Remote Control Commands
R&S
®
RTB2000
286User Manual 1333.1611.02 ─ 03
Parameters:
<Range> WITHin | OUTSide | SHORter | LONGer
WITHin
Finds pulses inside the range width ± Δt.
OUTSide
Finds pulses outside the range width ± Δt.
SHORter
Finds pulses shorter than the given width.
LONGer
Finds pulses longer than the given width.
*RST: WITH
Example:
Chapter 15.2.3.1, "Searching for a Pulse of Specified Width",
on page 246
SEARch:TRIGger:WIDTh:WIDTh <Width>
Sets the reference pulse width, the nominal value for comparisons.
Parameters:
<Width> Default unit: s
Example:
Chapter 15.2.3.1, "Searching for a Pulse of Specified Width",
on page 246
SEARch:TRIGger:WIDTh:DELTa <DeltaWidth>
Sets a range Δt to the reference pulse width set with
SEARch:TRIGger:WIDTh:
WIDTh
, if SEARch:TRIGger:WIDTh:RANGe is set to WITHin or OUTSide.
Parameters:
<DeltaWidth> Range: Lower limit depends on the resolution, practically no
upper limit
Example:
Chapter 15.2.3.1, "Searching for a Pulse of Specified Width",
on page 246
15.6.4.4 Peak Search Configuration
SEARch:MEASure:PEAK:POLarity
.................................................................................. 286
SEARch:MEASure:LEVel:PEAK:MAGNitude.....................................................................287
SEARch:MEASure:PEAK:POLarity <Polarity>
Indicates the polarity of a the pulse to be searched for a peak.
Parameters:
<Polarity> POSitive | NEGative | EITHer
*RST: POS
Waveform Analysis
Loading ...
Loading ...
Loading ...