Remote Control Commands
R&S
®
RTB2000
443User Manual 1333.1611.02 ─ 03
WGENerator:SWEep:TIME............................................................................................. 448
WGENerator:SWEep:TYPE............................................................................................ 448
WGENerator:SWEep[:ENABle]........................................................................................448
WGENerator:FUNCtion <Function>
Selects the function to be generated.
Parameters:
<Function> DC | SINusoid | SQUare | PULSe | TRIangle | RAMP | SINC |
ARBitrary | EXPonential
WGENerator:VOLTage <Amplitude>
Defines the amplitude value (peak-to-peak value) of the selected generator function.
Parameters:
<Amplitude> Numeric value
Range: 6.0000E-02 to 6.00000E+00
*RST: 5.0000E-01
Default unit: Vpp
WGENerator:VOLTage:OFFSet <Offset>
Sets the DC offset of the selected generator function.
Parameters:
<Offset> Numeric value
Range: -3.00000E+00 to 3.00000E+00
*RST: 0.00E+00
Default unit: V
WGENerator:FREQuency <Frequency>
Defines the frequency value of the selected generator function.
Parameters:
<Frequency> Numeric value depending on the selected generator function
Range: 1.000E-01 to 5.000000000E+4 (sine / square)
1.000E-01 to 1.000000000E+4 (pulse / triangle /
ramp)
*RST: 1.00000000E+3
WGENerator:FUNCtion:PULSe:DCYCle <DutyCycle>
Defines the duty cycle value of the generator function pulse.