Measurements
R&S
®
RTB2000
108User Manual 1333.1611.02 ─ 03
Figure 7-3: Cursor measurement with vertical and horizontal cursors and Set To Trace
Results = below the grid
Cursor lines 1, 2, 3 = not active
Cursor line 4 = active, can be moved by turning the NAVIGATION knob
To configure cursor measurements
1. Press the CURSOR key.
The cursors are activated with the latest setting.
2. Press the CURSOR key again.
The "Cursor" menu opens.
3. Select the cursor "Type".
4. Select the "Source", the waveform you want to measure.
5. Set additional settings if necessary:
Track Scaling, Coupling, Set To Trace, or Set
To Screen.
6. To change the position of a cursor line, you can use several methods:
Drag the cursor line on the screen.
Press the NAVIGATION knob repeatedly until the required cursor line is active
(marked with a solid line).
Turn the knob to move the line.
Tap the corresponding result value in the result line at the bottom.
The keypad opens, and you can enter an exact value.