Measurements
R&S
®
RTB2000
100User Manual 1333.1611.02 ─ 03
Quick measurement is not available on math and reference waveforms. Channels
other than the selected one are switched off in quick measurement mode. When quick
measurement is active, cursor measurements are not possible, but you can use auto-
matic measurements in parallel.
► Press the QUICKMEAS key again to deactivate quick measurement.
The results are deleted on the display.
Remote commands:
●
MEASurement<m>:AON on page 306
●
MEASurement<m>:AOFF on page 306
●
MEASurement<m>:ALL[:STATe] on page 306
●
MEASurement<m>:ARESult? on page 306
7.2 Automatic Measurements
You can perform up to four different measurements simultaneously.
To configure automatic measurements in the Measure menu
1. Press the MEAS key.
2. In the menu, select the "Meas. Place", the number of the measurement that you
want to configure.
3. If the measurement is off, enable "Measure <n>".
4. Select the measurement type:
a) Tap "Type"
b) Select the tab of the required measurement category.
c) Select the measurement type.
The measurement types are described in
Chapter 7.2.2, "Measurement Types",
on page 101.
5. Select the "Source".
The selection list shows all possible sources. If the waveform is not active, it is acti-
vated automatically when selected as measurement source.
6. Some measurement types require additional settings. Scroll down the menu, and
adjust the additional settings if necessary.
See also:
Chapter 7.2.3, "Settings for Automatic Measurements", on page 105.
7.2.1 Measurement Results
The measurement results are shown in a line below the grid.
Automatic Measurements