Remote Control Commands
R&S
®
RTB2000
348User Manual 1333.1611.02 ─ 03
Suffix:
<m>
.
1..4
Return values:
<Xincrement> Time in s
Example: See
Chapter 15.2.1.4, "Reading Waveform Data in Unsigned
Integer Format"
, on page 243
Usage: Query only
CHANnel<m>:DATA:YORigin?
CHANnel<m>:DATA:ENVelope:YORigin?
CALCulate:QMATh:DATA:YORigin?
MASK:DATA:YORigin?
LOGic<p>:DATA:YORigin?
DIGital<m>:DATA:YORigin?
REFCurve<m>:DATA:Yorigin?
Return the voltage value for binary value 0 of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined
(
FORM UINT, 8|16|32).
Suffix:
<m>
.
1..4
Return values:
<Yorigin> Voltage in V
Example: See
Chapter 15.2.1.4, "Reading Waveform Data in Unsigned
Integer Format"
, on page 243
Usage: Query only
CHANnel<m>:DATA:YINCrement?
CHANnel<m>:DATA:ENVelope:YINCrement?
CALCulate:QMATh:DATA:YINCrement?
MASK:DATA:YINCrement?
LOGic<p>:DATA:YINCrement?
DIGital<m>:DATA:YINCrement?
REFCurve<m>:DATA:YINCrement?
Return the voltage value per bit of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined
(
FORM UINT, 8|16|32).
Suffix:
<m>
.
1..4
Return values:
<Yincrement> Voltage in V