Remote Control Commands
R&S
®
RTB2000
345User Manual 1333.1611.02 ─ 03
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Return values:
<Data> List of values according to the format settings.
Usage: Query only
REFCurve<m>:DATA:HEADer?
Returns information on the reference waveform.
Table 15-4: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in Samples 200000
4 Number of values per sample interval, usually 1. 1
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Parameters:
<Header> Comma-separated value list
Example: -9.477E-008,9.477E-008,200000,1
Usage: Query only
15.9.1.4 Masks
In addition to the commands described below, consider also the following commands:
MASK:DATA:XINCrement? on page 347
MASK:DATA:XORigin? on page 347
MASK:DATA:YINCrement? on page 348
MASK:DATA:YORigin? on page 348
MASK:DATA:YRESolution? on page 349
MASK:DATA?
Returns the data of the mask. The mask consists of two limit curves.
To set the export format, use
FORMat[:DATA] on page 339.
Return values:
<Data> List of values according to the format settings - the y-values of
the mask points. The list contains two values for each sample
interval.
Documenting Results