Remote Control Commands
R&S
®
RTB2000
319User Manual 1333.1611.02 ─ 03
Value Description Queries for results
HORizontal Sets two horizontal cursor lines and measures the
voltages at the two cursor positions and the delta
of the two values.
CURSor<m>:Y1Position
CURSor<m>:Y2Position
CURSor<m>:YDELta[:VALue]?
CURSor<m>:YDELta:SLOPe?
VERTical Sets two vertical cursor lines and measures the
time from the trigger point to each cursor, the time
between the two cursors and the frequency calcu-
lated from that time.
CURSor<m>:X1Position
CURSor<m>:X2Position
CURSor<m>:XDELta[:VALue]?
CURSor<m>:XDELta:INVerse?
HVERtical Combines the HORizontal cursor and VERTical
cursor measurements. Two horizontal and two
vertical cursor lines are set and the voltages and
time from the trigger point are measured at the
cursor positions, as well as the delta of the voltage
and time values.
CURSor<m>:Y1Position
CURSor<m>:Y2Position
CURSor<m>:YDELta[:VALue]?
CURSor<m>:X1Position
CURSor<m>:X2Position
CURSor<m>:XDELta[:VALue]?
CURSor<m>:TRACking[:STATe] <State>
If set to ON, the V-Marker cursor measurement is enabled.
Suffix:
<m>
.
1, the suffix is irrelevant.
Parameters:
<State> ON | OFF
*RST: OFF
CURSor<m>:X1Position <Xposition1>
CURSor<m>:X2Position <Xposition2>
The commands specify or return the positions of vertical cursor lines on the x-axis
(time, frequency for FFT).
Suffix:
<m>
.
1, the suffix is irrelevant.
Parameters:
<Xposition1>
<Xposition2>
Range: Depends on the current instrument settings, for
example, horizontal position.
Default unit: s
CURSor<m>:Y1Position <Yposition1>
CURSor<m>:Y2Position <Yposition2>
The commands specify or return the positions of horizontal cursor lines on the y-axis
(voltage, current, level for FFT).
Measurements