Trigger
R&S
®
RTB2000
65User Manual 1333.1611.02 ─ 03
TRIGGER > "Trigger Type" = "Pattern" > "Edit Pattern"
Figure 5-7: Pattern trigger with logic editor
H | L | X, Set All.............................................................................................................65
And | Or.........................................................................................................................65
Duration.........................................................................................................................66
True | False................................................................................................................... 66
Time limitation...............................................................................................................66
H | L | X, Set All
Defines the pattern by selecting the state "H" (high), "L" (low) or "X" (do not care) for
each active analog and digital channel.
The word length of the pattern depends on the number of available analog and digital
channels.
Analog channels: 2 bit for 2-channel instruments, 4 bit for 4-channel instruments.
Digital (16 bit): the logic channels D0, D1,...,D15 are only available with MSO option
R&S RTB-B1.
Thus the pattern can have 2, 4, 18, or 20 bits.
Use "Set All" to set all channels to the same state.
Remote command:
TRIGger:A:PATTern:SOURce on page 273
And | Or
Sets the logical combination of the channel states.
"AND"
All defined states must be true.
"OR"
At least one of the defined states must be true.
Remote command:
TRIGger:A:PATTern:FUNCtion on page 273