Rohde & Schwarz RTB2002 Oscilloscope

User Manual - Page 134

For RTB2002.

PDF File Manual, 510 pages, Read Online | Download pdf file

RTB2002 photo
Loading ...
Loading ...
Loading ...
Applications
R&S
®
RTB2000
126User Manual 1333.1611.02 ─ 03
Remote command:
DISPlay:XY:Y2Source on page 335
8.4 Digital Voltmeter
The integrated three-digit digital voltmeter simplifies measurements, in particular for
service personnel. You can measure up to four parameters on different sources at the
same time and define the position of the measurement results.
The following DVM measurements are available:
DC: mean value of the signal
AC+DC RMS: RMS value of the signal
AC RMS: RMS value of the signal's AC component
The digital voltmeter captures input data with the selected vertical sensitivity and the
basic accuracy of the ADC. It is independent from the capture settings and the post
processing. All measurements are based on a measurement interval, which ensures
reliable results over the defined frequency range.
Figure 8-1: Results of 4 meter measurements. The measurement source is indicated by the channel
color. The source of meter 1 is channel 1, and it is clipped.
Remote commands to read meter results:
DVM<m>:RESult[:ACTual]? on page 336
DVM<m>:RESult[:ACTual]:STATus? on page 337
8.4.1 Using the Meter
To activate meter measurements
Use one of the following ways:
Tap the "Meter" icon on the toolbar.
Press the
APPS SELECTION key.
Tap "Meter".
Digital Voltmeter
Loading ...