Rohde & Schwarz RTB2002 Oscilloscope

User Manual - Page 221

For RTB2002.

PDF File Manual, 510 pages, Read Online | Download pdf file

RTB2002 photo
Loading ...
Loading ...
Loading ...
Serial Bus Analysis
R&S
®
RTB2000
213User Manual 1333.1611.02 ─ 03
Identifier ← Identifier condition
Defines the identifier pattern in binary and hexadecimal format. To set an individual
binary bit or hex nibble (half byte), tap it and enter it via the on-screen keypad.
"Bin"
String containing the binary pattern with a maximum of 64 bits. Char-
acters 0, 1 and X are allowed.
"Hex"
String containing the hexadecimal pattern with a maximum of 8 bytes.
Characters 0 to F and X are allowed.
Remote command:
TRIGger:A:LIN:IDENtifier on page 421
Compare ← Identifier condition
Sets the identifier comparison condition: If the identifier pattern contains at least one X
(don't care), you can trigger on values equal or not equal to the specified value. If the
pattern contains only 0 and 1, you can also trigger on a range greater than or lower
than the specified value.
Remote command:
TRIGger:A:LIN:ICONdition on page 421
Symbolic ID ← Identifier condition
If a label list with symbolic names was loaded and applied in the bus configuration, you
can select a symbolic name from the list instead of entering the numeric identifier. The
instrument triggers on the identifier of the selected node.
Data condition
The data condition consists of the following settings:
Length of the data pattern
Comparison
Data pattern
Data ← Data condition
Defines the length of the data pattern - the number of bytes in the pattern.
Remote command:
TRIGger:A:LIN:DLENgth on page 422
Compare ← Data condition
Sets the comparison condition: If the pattern contains at least one X (don't care), you
can trigger on values equal or not equal to the specified value. If the pattern contains
only 0 and 1, you can also trigger on a range greater than or lower than the specified
value.
Remote command:
TRIGger:A:LIN:DCONdition on page 422
Bin / Hex ← Data condition
Specifies the data pattern to be triggered. To set an individual binary bit or hexadeci-
mal nibble (half byte), tap it and enter it via the on-screen keypad. Make sure to specify
complete bytes.