Rohde & Schwarz RTB2002 Oscilloscope

User Manual - Page 445

For RTB2002.

PDF File Manual, 510 pages, Read Online | Download pdf file

RTB2002 photo
Loading ...
Loading ...
Loading ...
Remote Control Commands
R&S
®
RTB2000
437User Manual 1333.1611.02 ─ 03
Depending on the current settings, the memory can contain more data samples than
the screen is able to display. In this case, you can decide which data will be saved:
samples stored in the memory or only the displayed samples.
Note: The sample range can be changed only in STOP mode. If the acquisition is run-
ning, DEF is always used automatically. If the acquisition has been stopped, data can
be read from the memory, and all settings are available.
Suffix:
<m>
.
0..15
Selects the logic channel
<p> 1..2
Selects the logic pod
Setting parameters:
<PointSelection> DEFault | MAXimum | DMAXimum
Sets the range for data queries.
DEFault
Waveform points that are visible on the screen. At maximum
waveform rate, the instrument stores more samples than visible
on the screen, and DEF returns less values than acquired.
MAXimum
All waveform samples that are stored in the memory. Only avail-
able if acquisition is stopped.
DMAXimum
Display maximum: Waveform samples stored in the current
waveform record but only for the displayed time range. At maxi-
mum waveform rate, the instrument stores more samples than
visible on the screen, and DMAX returns more values than DEF.
Only available if acquisition is stopped.
*RST: DEFault
Return values:
<Points> Number of data points in the selected range.
Default unit: Samples
See also:
CHANnel<m>:DATA:POINts
15.12.2 Parallel Buses
Parallel Bus - Line Configuration...........................................................................437
Parallel Clocked Bus Configuration.......................................................................439
Parallel Buses - Decode Results...........................................................................440
15.12.2.1 Parallel Bus - Line Configuration
BUS<b>:PARallel:WIDTh <BusWidth>
Sets the number of lines to be analyzed for the parallel bus.
Mixed Signal Option (Option R&S RTB-B1)
Loading ...
Loading ...
Loading ...