Rohde & Schwarz RTB2002 Oscilloscope

User Manual - Page 125

For RTB2002.

PDF File Manual, 510 pages, Read Online | Download pdf file

RTB2002 photo
Loading ...
Loading ...
Loading ...
Applications
R&S
®
RTB2000
117User Manual 1333.1611.02 ─ 03
Test
Performs a mask test for the selected signal, i.e. the signal amplitudes are compared
with the specified mask. If the amplitude exceeds the limits of the mask, a violation is
detected.
Remote command:
MASK:TEST on page 326
Copy Channel
Creates a mask from the envelope waveform of the selected channel and stores it in
the instrument.
Remote command:
MASK:SOURce on page 323
MASK:CHCopy on page 323
Y-Position
Moves the mask vertically within the display. The current position is given in divisions.
Remote command:
MASK:YPOSition on page 323
Stretch Y
Changes the vertical scaling to stretch or compress the mask in y-direction.
Remote command:
MASK:YSCale on page 324
Width X
Changes the width of the mask in horizontal direction. The specified factor in divisions
is added to the positive x-values and subtracted from the negative x-values of the
mask limits in relation to the mask center. Thus, the left half of the mask is pulled to the
left, the right half is pulled to the right.
Remote command:
MASK:XWIDth on page 324
Width Y
Changes the width of the mask in vertical direction. The specified number of divisions
is added to the y-values of the upper mask limit and subtracted from the y-values of the
lower mask limit. Thus, the upper half of the mask is pulled upwards, the lower half is
pulled down, and the overall height of the mask is twice the "Width Y".
Remote command:
MASK:YWIDth on page 324
Save
Saves the mask in an instrument-specific format (MSK). The complete mask definition
- envelope waveform with width, stretch and position settings - is stored.
Remote command:
MASK:SAVE on page 324 and commands described in Chapter 15.9.1.4, "Masks",
on page 345.
Mask Testing
Loading ...
Loading ...
Loading ...