Remote Control Commands
R&S
®
RTB2000
292User Manual 1333.1611.02 ─ 03
SEARch:TRIGger:PATTern:WIDTh:RANGe......................................................................293
SEARch:TRIGger:PATTern:WIDTh[:WIDTh]..................................................................... 293
SEARch:TRIGger:PATTern:WIDTh:DELTa.......................................................................293
SEARch:TRIGger:PATTern:SOURce <Pattern>
Specifies the search pattern - the state for each channel.
Parameters:
<Pattern> String parameter
String containing 0, 1, X|x for each channel. The order of chan-
nels is fixed: CH1 CH2 [CH3 CH4] D0 D1 D2 D3 ...
Example:
SEAR:TRIG:PATT:SOUR '1X10'
CH1, CH3 are high, CH4 is low. These states are logically com-
bined with
SEARch:TRIGger:PATTern:FUNCtion. CH2 does
not matter (don't care).
SEARch:TRIGger:PATTern:FUNCtion <Function>
Sets the logical combination of the channel states.
Parameters:
<Function> AND | OR | NAND | NOR
AND
The required states of all channels must appear in the input sig-
nal at the same time.
OR
At least one of the channels must have the required state.
NAND
"Not and" operator, at least one of the channels does not have
the required state.
NOR
"Not or" operator, none of the channels has the required state.
*RST: AND
SEARch:TRIGger:PATTern:LEVel<n> <ThresholdLevel>
Sets the threshold value for the specified source channel. You can set different levels
for the channels
Suffix:
<n>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ThresholdLevel> Range: Depends on vertical scale