Remote Control Commands
R&S
®
RTB2000
336User Manual 1333.1611.02 ─ 03
DVM<m>:ENABle.......................................................................................................... 336
DVM<m>:SOURce.........................................................................................................336
DVM<m>:TYPE............................................................................................................. 336
DVM<m>:RESult[:ACTual]?............................................................................................ 336
DVM<m>:RESult[:ACTual]:STATus?................................................................................337
DVM<m>:ENABle <VoltmeterEnable>
Enables and disables all configured meter measurements.
Suffix:
<m>
.
1..4
The suffix is irrelevant.
Parameters:
<VoltmeterEnable> ON | OFF
*RST: OFF
DVM<m>:SOURce <Source>
Selects an analog channel as the source of the selected meter measurement.
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4
CH3 and CH4 are only available with 4-channel instruments.
DVM<m>:TYPE <MeasurementType>
Sets the measurement type for the indicated DVM measurement.
Set OFF to disable the measurement.
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Parameters:
<MeasurementType> DC | ACDCrms | ACRMs | OFF
DC: mean value of the signal
ACDCrms: RMS value of the signal (AC+DC RMS)
ACRMs: RMS value of the signal's AC component (AC RMS)
OFF: disables the selected meter measurement.
*RST: OFF
DVM<m>:RESult[:ACTual]?
Returns the current value of the selected meter measurement.