Remote Control Commands
R&S
®
RTB2000
421User Manual 1333.1611.02 ─ 03
TRIGger:A:LIN:CHKSerror <ChecksumError>
Triggers on a checksum error. The checksum verifies the correct data transmission. It
is the last byte of the frame response. The checksum includes not only the data but
also the protected identifier (PID).
The command is relevant if
TRIGger:A:LIN:TYPE is set to ERRCondition.
Parameters:
<ChecksumError> ON | OFF
*RST: ON
TRIGger:A:LIN:IPERror <IdParityError>
Triggers on a parity error. Parity bits are the bits 6 and 7 of the identifier. They verify
the correct transmission of the identifier.
The command is relevant if
TRIGger:A:LIN:TYPE is set to ERRCondition.
Parameters:
<IdParityError> ON | OFF
*RST: OFF
TRIGger:A:LIN:SYERror <SyncError>
Triggers if synchronization caused an error.
The command is relevant if
TRIGger:A:LIN:TYPE is set to ERRCondition.
Parameters:
<SyncError> ON | OFF
*RST: OFF
TRIGger:A:LIN:ICONdition <IdentifierCondition>
Sets the comparison condition for the identifier: If the pattern contains at least one X
(don't care), you can trigger on values equal or not equal to the specified value. If the
pattern contains only 0 and 1, you can also trigger on a range greater than or lower
than the specified value.
The command is relevant if
TRIGger:A:LIN:TYPE is set to ID or IDDT.
Parameters:
<IdentifierCondition> EQUal | NEQual | GTHan | LTHan
*RST: EQ
TRIGger:A:LIN:IDENtifier <Identifier>
Defines the identifier pattern.
The command is relevant if TRIGger:A:LIN:TYPE is set to ID or IDDT.