Status Reporting System
R&S
®
RTB2000
481User Manual 1333.1611.02 ─ 03
Bit No. Meaning
8
NOALigndata
This bit is set if no alignment data is available - the instrument is uncalibrated.
9
LIMit
This bit is set if a limit value is violated (see
Chapter C.3.4.3, "STATus:QUEStionable:LIMit reg-
ister", on page 481).
10 to 11 Not used
12
MASK
This bit is set if a mask value is violated (see
Chapter C.3.4.4, "STATus:QUEStionable:MASK
register", on page 482
13 to 14 Not used
15 This bit is always 0.
C.3.4.1 STATus:QUEStionable:COVerload register
This register contains all information about overload of the channels. The bit is set if
the assigned channel is overloaded.
Table C-5: Bits in the STATus:QUEStionable:COVerload register
Bit No. Meaning
0
CHANnel1
1
CHANnel2
2
CHANnel3
3
CHANnel4
C.3.4.2 STATus:QUEStionable:TEMPerature register
This register contains information about the instrument's temperature.
Table C-6: Bits in the STATus:QUEStionable:TEMPerature register
Bit No. Meaning
0
TEMP WARN
This bit is set if a temperature warning on channel 1, 2, 3 or 4 occured.
1
TEMP ERRor
This bit is set if a temperature error on channel 1, 2, 3 or 4 occured.
C.3.4.3 STATus:QUEStionable:LIMit register
This register contains information about the observance of the limits of measurements.
This bit is set if the limits of the main or additional measurement of the assigned mea-
surement are violated.
Contents of the Status Registers
Loading ...
Loading ...