Remote Control Commands
R&S
®
RTB2000
314User Manual 1333.1611.02 ─ 03
Usage: Query only
MEASurement<m>:STATistics:VALue<n>?
Returns the n
th
statistical value from the indicated buffer place.
Note: Valid buffered values can only be read if the acquisition is stopped. As long as
the acquisition is running, the buffer contents is changing and the buffered values are
not valid for reading.
Suffix:
<m>
.
1..4
Selects the measurement place.
<n> *
Buffer place. The buffer size is limited by MEASurement<m>:
STATistics:WEIGht
.
Return values:
<StatisticValue> Statistical value
Usage: Query only
EXPort:MEASurement<m>:STATistics:NAME
Defines the path and filename of the statistics file. The file format is CSV. If the file
already exists, it will be overwritten without notice.
Suffix:
<m>
.
1..4
Selects the measurement place.
Parameters:
<FileName> String parameter
EXPort:MEASurement<m>:STATistics:SAVE
Saves statistical results of the indicated measurement place to the file that is defined
by the EXPort:MEASurement<m>:STATistics:NAME command.
Suffix:
<m>
.
1..4
Selects the measurement place.
Usage: Event
See also:
EXPort:MEASurement:STATistics:ALL:SAVE on page 315.
EXPort:MEASurement:STATistics:ALL:NAME
Defines the path and filename of the statistics file. The file format is CSV. If the file
already exists, it will be overwritten.