Remote Control Commands
R&S
®
RTB2000
275User Manual 1333.1611.02 ─ 03
TRIGger:A:PATTern:WIDTh[:WIDTh] <PatternWidth>
For the ranges WITHin and OUTSide, the <PatternWidth> defines the center of a
range which is defined by the limits ±<Delta>.
For the ranges SHORter and LONGer, the pattern width defines the maximum and
minimum values, respectively.
Parameters:
<PatternWidth> Default unit: s
TRIGger:A:PATTern:WIDTh:DELTa <PatternDelta>
Defines a range around the pattern width value specified using
TRIGger:A:
PATTern:WIDTh[:WIDTh]
.
Parameters:
<PatternDelta> Default unit: s
15.5.6 Timeout Trigger
To set the trigger level (threshold), use:
TRIGger:A:LEVel<n>[:VALue] on page 268
CHANnel<m>:THReshold on page 256
TRIGger:A:FINDlevel on page 268
CHANnel<m>:THReshold:FINDlevel on page 256
CHANnel<m>:THReshold:HYSTeresis on page 257
TRIGger:A:TIMeout:RANGe............................................................................................275
TRIGger:A:TIMeout:TIME............................................................................................... 275
TRIGger:A:TIMeout:RANGe <Range>
Sets the relation of the signal level to the threshold.
Parameters:
<Range> HIGH | LOW | EITHer
HIGH
The signal level stays above the trigger level.
LOW
The signal level stays below the trigger level.
EITHer
The signal level stays above or below the trigger level.
*RST: HIGH
TRIGger:A:TIMeout:TIME <Time>
Sets the time limit for the timeout at which the instrument triggers.