General Instrument Setup
R&S
®
RTB2000
146User Manual 1333.1611.02 ─ 03
To remove all waveforms and results from the display, press the CLEAR SCREEN
key at the front panel.
Persistence
Defines the persistence (afterglow effect) of the waveform on the display.
"Off"
Deactivates persistence.
"Manual"
User-defined persistence according to "Time" setting.
"Infinite"
Activates persistence with infinite duration. Each new data point
remains on the screen infinitely until this setting is changed or the
persistence is cleared
Remote command:
DISPlay:PERSistence:TYPE on page 360
Time ← Persistence
Specifies the afterglow duration if "Persistence" is set to "Manual". Each new data
point remains on the screen for the time defined here. Values range from 50 ms (=
default) to 12.8 s.
Remote command:
DISPlay:PERSistence:TIME on page 360
Intensities
Provides functions to define the brightness (relative luminous intensity) of display ele-
ments and control LEDs.
You can also open this menu directly using the INTENSITY key.
Waveform ← Intensities
Defines the brightness of the waveform lines in the diagram. Enter a percentage
between 0 (barely visible) and 100% or turn the NAVIGATION knob to adjust the wave-
form brightness directly. The default value is 50%.
Remote command:
DISPlay:INTensity:WAVeform on page 362
Grid ← Intensities
Defines the brightness of the grid lines in the diagram. Enter a percentage between 0
(barely visible) and 100% or turn the NAVIGATION knob to adjust the grid brightness
directly. The default value is 50%.
Display Settings