Remote Control Commands
R&S
®
RTB2000
268User Manual 1333.1611.02 ─ 03
TRIGger:A:HOLDoff:TIME <HoldOffTime>
Defines the holdoff time. The next trigger occurs only after the holdoff time has passed.
Parameters:
<HoldOffTime> Default unit: s
15.5.2 Edge Trigger
TRIGger:A:EDGE:SLOPe............................................................................................... 268
TRIGger:A:LEVel<n>[:VALue]......................................................................................... 268
TRIGger:A:FINDlevel......................................................................................................268
TRIGger:A:EDGE:COUPling........................................................................................... 269
TRIGger:A:EDGE:FILTer:HFReject..................................................................................269
TRIGger:A:EDGE:FILTer:NREJect...................................................................................269
TRIGger:A:EDGE:SLOPe <Slope>
Sets the slope for the edge trigger.
Parameters:
<Slope> POSitive | NEGative | EITHer
POSitive
Rising edge, a positive voltage change
NEGative
Falling edge, a negative voltage change
EITHer
The rising as well as the falling edge
*RST: POSitive
TRIGger:A:LEVel<n>[:VALue] <Level>
Sets the trigger threshold voltage for edge, width, and timeout trigger.
Suffix:
<n>
.
1..5
Selects the trigger input. 1...4 select the corresponding analog
channel, 5 is the external trigger input. The number of channels
depends on the instrument.
Parameters:
<Level> Range: Depends on vertical scale.
Default unit: V
TRIGger:A:FINDlevel
Sets the trigger level to 50% of the signal amplitude.
Usage: Event
Trigger