General Instrument Setup
R&S
®
RTB2000
151User Manual 1333.1611.02 ─ 03
ance of the probe-oscilloscope system and introduces measurement errors resulting in
distorted waveforms and inaccurate results.
Two connector pins for compensation are located at the front panel. The left pin is on
ground level. The next pin supplies a square wave signal for the adjustment.
1.
Press the
APPS SELECTION key.
2. Tap "Probe Adjust".
3. Follow the instructions of the wizard. It guides you through the compensation proc-
ess.
Use the compensation trimmer of the probe to get optimum square wave response.
For details, refer to the documentation of your probe.
10.8 Options
All options are activated by license keys. No additional installation or hardware change
is required.
Unregistered licenses
Unregistered licenses are not assigned to a particular instrument. The instrument
accepts only registered licenses. If your license is delivered unregistered, use the
online tool R&S License Manager to register the license for your instrument. The regis-
tration of a permanent license is irreversible, so ensure that you register it for the cor-
rect instrument. The address of the tool is
https://extranet.rohde-schwarz.com/service.
The "Active Options" tab provides information on installed software options. Here you
can install new options or deactivate existing options using license keys.
The "Inactive Options" tab lists all deactivated and expired options.
The "Deactivated Options" tab shows all deactivated options with their deactivation
information and provides a function to export the deactivation response. The response
is required by the R&S License Manager.