General Instrument Setup
R&S
®
RTB2000
148User Manual 1333.1611.02 ─ 03
Track Grid ← Grid
If enabled, the grid moves with the waveforms, if you change the waveform's position
in horizontal or vertical direction.
If disabled (= default), the grid remains centered on the display, if you change the
waveform's position.
10.3 Resetting the Instrument
Reset is helpful if the instrument is in undefined condition and cannot be operated.
To reset all waveform and measurement settings:
► Press Preset.
Preset does not change the display settings, for example, intensities and brightness.
To reset these settings, restore the factory settings.
To restore the factory settings:
1. Press Save Load.
2. Tap "Setup".
3. Tap "Factory Default".
10.4 Locking the Touchscreen
The Touch Lock key locks the touchscreen to prevent unintended use. When the
touchscreen is off, the key is illuminated. Press again to unlock the touchscreen.
10.5 Performing a Self-Alignment
The self-alignment aligns the data from several input channels vertically and horizon-
tally to synchronize the timebases, amplitudes and positions.
Recommendation on performing the self-alignment:
●
When putting the instrument into operation for the first time
●
After a firmware update
●
Once a week
●
When major temperature changes occur (> 5°)
Performing a Self-Alignment