Remote Control Commands
R&S
®
RTB2000
350User Manual 1333.1611.02 ─ 03
Parameters:
<WaveformSource> CH1..4 | D70 | D158 | MA1 | RE1..4
CH1..4
Analog channels CH1 | CH2 | CH3 | CH4
D70
Pod 1, digital channels D0 to D7 are exported together
D158
Pod 2, digital channels D8 to D15 are exported together.
MA1
Mathematic waveform
RE1..4
Reference waveforms RE1 | RE2 | RE3 | RE4
EXPort:WAVeform:NAME <FileName>
Defines the path and filename for a waveform data file that will be saved with
EXPort:
WAVeform:SAVE
. The data format and file extension is defined using FORMat[:
DATA]
.
Existing files will be overwritten.
You can change the storage location, file name and/or file format manually in the FILE
> "Waveforms" menu. Remote control uses the recent settings.
Parameters:
<FileName> String parameter
Example:
FORMAT CSV
EXPort:WAVeform:NAME "/USB_FRONT/WAVEFORMS/WFM01"
EXPort:WAVeform:SAVE
The waveform data is saved to WFM01.CSV.
EXPort:WAVeform:SAVE
Executes saving a waveform, for which the path and filename have been defined by
EXPort:WAVeform:NAME.
Usage: Event
15.9.3 Screenshots
This chapter describes remote commands used to save screenshots.
MMEMory:NAME........................................................................................................... 351
HCOPy[:IMMediate]....................................................................................................... 351
HCOPy:DATA?..............................................................................................................351
HCOPy:FORMat............................................................................................................ 351
HCOPy:LANGuage........................................................................................................ 351