Remote Control Commands
R&S
®
RTB2000
344User Manual 1333.1611.02 ─ 03
●
CALCulate:QMATh:DATA:YRESolution? on page 349
CALCulate:QMATh:DATA?
Returns the data of the math waveform points for transmission from the instrument to
the controlling computer. The waveforms data can be used in MATHLAB, for example.
To set the export format, use FORMat[:DATA] on page 339.
Return values:
<Data> List of values according to the format settings - voltages, or
magnitudes of a spectrum.
Usage: Query only
CALCulate:QMATh:DATA:HEADer?
Returns information on the math waveform.
Table 15-3: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in Samples 120000
4 Number of values per sample interval, usually 1. 1
Parameters:
<Header> Comma-separated value list
Example: -9.477E-008,9.477E-008,120000,1
Usage: Query only
15.9.1.3 Reference Waveforms
In addition to the commands described below, consider also the following commands:
●
REFCurve<m>:DATA:XINCrement? on page 347
●
REFCurve<m>:DATA:Xorigin? on page 347
●
REFCurve<m>:DATA:YINCrement? on page 348
●
REFCurve<m>:DATA:Yorigin? on page 348
●
REFCurve<m>:DATA:YRESolution? on page 349
REFCurve<m>:DATA?
Returns the data of the reference waveform for transmission from the instrument to the
controlling computer. The waveforms data can be used in MATLAB, for example.
To set the export format, use FORMat[:DATA] on page 339.