Remote Control Commands
R&S
®
RTB2000
370User Manual 1333.1611.02 ─ 03
Parameters:
<OutputMode> OFF | TRIGger | REFerence | MASK | GENerator
OFF
No output
TRIGger
Outputs a pulse when the instrument triggers.
Outputs a 10 MHz reference frequency.
MASK
Outputs a pulse when a mask is violated. This function is only
available if a mask is specified.
GENerator
Outputs the waveform specified using the function generator
(requires option R&S RTB-B6)
*RST: OFF
TRIGger:OUT:PLENgth <PulseLength>
Defines the pulse width of the pulse at the AUX OUT front connector (at trigger event
or mask viaolation).
Parameters:
<PulseLength> *RST: 1E-6
TRIGger:OUT:POLarity <Polarity>
Defines the polarity of the pulse at the AUX OUT front connector (at trigger event or
mask viaolation).
Parameters:
<Polarity> POSitive | NEGative
*RST: POS
15.11 Serial Bus Analysis
●
General................................................................................................................. 371
●
SPI and SSPI (Option R&S RTB-K1)....................................................................373
●
I²C......................................................................................................................... 384
●
UART (Option R&S RTB-K2)................................................................................ 395
●
CAN.......................................................................................................................402
●
LIN (Option R&S RTB-K3).................................................................................... 418
Serial Bus Analysis