Measurements
R&S
®
RTB2000
110User Manual 1333.1611.02 ─ 03
"Vertical &
Horiz."
Combines the "Horizontal" cursor and "Vertical" cursor measure-
ments. Two horizontal and two vertical cursor lines are set. The vol-
tages and times are measured at the cursor positions, as well as the
delta of the voltage and time values.
Results: t1, t2, Δt, V1, V2, ΔV
"V-Marker"
Sets two vertical cursors and measures the values of the waveform at
the crossing points of the cursor lines and the waveform. Also, the dif-
ferences of the two values in x- and y-direction are displayed.
Results: t1, V1, t2, V2, Δt, ΔV
Remote command:
CURSor<m>:FUNCtion on page 318
CURSor<m>:TRACking[:STATe] on page 319 (V-Marker)
Source
Defines the source of the cursor measurement as one of the active waveforms.
You can use cursors on analog input signals, math waveform, reference waveforms,
XY-diagram, and FFT waveform.
If option R&S RTB -B1 is installed, you can use the vertical cursor to measure individ-
ual logic channels, and the V-Marker to measure pods.
Remote command:
CURSor<m>:SOURce on page 318
Track Scaling
If enabled, the cursor lines are adjusted when the vertical or horizontal scales are
changed. The cursor lines keep their relative position to the waveform.
If disabled, the cursor lines remain on their position on the display if the scaling is
changed.
Remote command:
CURSor<m>:TRACking:SCALe[:STATe] on page 320
Coupling
If enabled, the cursors lines are coupled and moved together.
Press the NAVIGATION key to select whether both cursors or one cursor is moved. If
coupling is disabled, pressing the NAVIGATION key toggles the single cursor lines.
Remote command:
CURSor<m>:XCOupling on page 320
CURSor<m>:YCOupling on page 320
Set To Trace
Autoset for cursor lines, sets the cursor lines to typical points of the waveform depend-
ing on the selected cursor type. For example, for voltage measurement ("Horizontal"),
the cursor lines are set to the upper and lower peaks of the waveform. For time mea-
surement ("Vertical"), the cursor lines are set to the edges of two consecutive positive
or two consecutive negative pulses.
Remote command:
CURSor<m>:SWAVe on page 320