Remote Control Commands
R&S
®
RTB2000
351User Manual 1333.1611.02 ─ 03
HCOPy:SIZE:X?............................................................................................................ 352
HCOPy:SIZE:Y?............................................................................................................ 352
HCOPy:COLor:SCHeme.................................................................................................352
MMEMory:NAME <file_name>
Defines the file name to store an image of the display with
HCOPy[:IMMediate].
Parameters:
<file_name> String parameter
Example: see Chapter 15.2.1.1, "Saving Screenshots to File", on page 241
HCOPy[:IMMediate]
Saves a screenshot to external memory.
Before starting, make sure that:
The path for storage is defined correctly by
MMEMory:CDIRectory
The file name for storage is defined by
MMEMory:NAME.
Example: see
Chapter 15.2.1.1, "Saving Screenshots to File", on page 241
Usage: Event
HCOPy:DATA?
Returns the data of the image file. The file format is defined using
HCOPy:LANGuage
(BMP | PNG)
Return values:
<ScreenShot> 488.2 block data
Usage: Query only
HCOPy:FORMat <Format>
HCOPy:LANGuage <Format>
Defines the format of the screenshot.
Parameters:
<Format> BMP | PNG
BMP | PNG | GIF
File formats for saved screenshots.
BMP: Windows Bitmap Format
PNG: Portable Network Graphic
*RST: PNG
Example: see
Chapter 15.2.1.1, "Saving Screenshots to File", on page 241