Remote Control Commands
R&S
®
RTB2000
261User Manual 1333.1611.02 ─ 03
Example:
ACQ:POIN:AUT ON
TIM:SCAL 1e-9
ACQ:POIN?;:SYST:ERR:ALL?
-> received 10000;0,"No error"
TIM:SCAL 5e-3
ACQ:POIN?;:SYST:ERR:ALL?
-> received 20000000;0,"No error"
ACQuire:POINts[:VALue]
Defines a record length value, the number of recorded waveform points in a segment.
The command turns
ACQuire:POINts:AUTomatic OFF.
If
ACQuire:POINts:AUTomatic is turned ON, the query ACQuire:POINts? returns
the automatically set record length.
Each predefined record length corresponds to a maximum number of history seg-
ments, which are stored in the instrument's memory. If option R&S RTB-K15 is instal-
led, you can display the history segments.
Available record length values are:
10 kSa (13107 history segments)
20 kSa (6553 history segments)
50 kSa (2621 history segments)
100 kSa (1456 history segments)
200 kSa (771 history segments)
500 kSa (319 history segments)
1 MSa (159 history segments)
2 MSa (80 history segments)
5 MSa (32 history segments)
10 MSa (16 history segments)
20 MSa (8 history segments)
Parameters:
<RecordLength> Record length in Sa.
If the entered value differs from the predefined values, the
instrument sets the closest value.
CHANnel<m>:TYPE <DecimationMode>
Selects the method to reduce the data stream of the ADC to a stream of waveform
points with lower sample rate.
Suffix:
<m>
.
1..4
The command affects all channels regardless of the indicated
channel number. The suffix can be omitted.