Trigger
R&S
®
RTB2000
62User Manual 1333.1611.02 ─ 03
"Inside", "Out-
side"
Triggers on pulses inside or outside a range specified with "Time t1"
and "Time t2".
This method is an alternative setting to the range definition with
"Time t" and "Variation". The values are interdependent. "Variation"
and "Time t" are adjusted, if you change t1 and t2, and vice versa.
Remote command:
TRIGger:A:WIDTh:RANGe on page 270
Time t
Sets the reference time, the nominal value for comparison settings "Width >",
"Width <", "Width =", and "Width".
Remote command:
TRIGger:A:WIDTh:WIDTh on page 270
Variation
Sets a range Δt to the reference "Time t", if comparison is set to "Width =" or "Width".
The instrument triggers on pulses inside or outside the range t±Δt.
Remote command:
TRIGger:A:WIDTh:DELTa on page 271
Time t1, Time t2
Set the lower and upper time limits defining the time range if "Width =" or "Width" is
set for comparison. "Time t" and "Variation" are adjusted accordingly.
Remote command:
TRIGger:A:WIDTh:RANGe on page 270
TRIGger:A:WIDTh:DELTa on page 271
Threshold
Threshold of the trigger source channel, used as trigger level for the width trigger.
See also "Threshold" on page 44 and "Trigger Level, Threshold" on page 58.
Remote command:
TRIGger:A:LEVel<n>[:VALue] on page 268
CHANnel<m>:THReshold on page 256
Hysteresis
Hysteresis of the trigger source channel, see
"Hysteresis" on page 44.
Remote command:
CHANnel<m>:THReshold:HYSTeresis on page 257
5.6 Video Trigger
The video or TV trigger is used to analyze analog baseband video signals. You can
trigger on baseband video signals from standard definition and high definition stand-
ards connected to an analog channel input or to the external trigger input.
The instrument triggers on the sync pulses.
Video Trigger
Loading ...
Loading ...
Loading ...