Select the conversion type to use for the probe on this channel. For TC probe
types, the available conversion types are:
Thermocouple Type
Volts
Refer to the Report of Calibration for the probe for information on the conver
-
sion type that the probe was calibrated for.
If the selected conversion type requires coefficients or other settings, the Con
-
version Settings button is enabled. Click the Conversion Settings button to dis
-
play the dialog for the selected conversion type. Otherwise, the Conversion
Settings button is disabled.
Note: The conversion settings are not written to the thermometer readout
until the Apply or OK button is clicked on this dialog!
Select the type of reference junction compensation to use on this channel using
the RJC type drop-down list. Available choices are Internal and External. If Ex-
ternal is chosen as the RJC type, the external RJC temperature must be entered
in the RJC temperature box.
Enter the external RJC temperature value in the RJC temperature box. This box
is disabled if the RJC type is set to Internal.
Note: The RJC temperature value must always be in °C!