FLUKE 9935-S LogWare II Software

User Manual for the Fluke 9935-S LogWare II Software - Page 118

For 9935-S.

PDF File Manual, 275 pages, Read Online | Download pdf file

9935-S photo
Loading ...
Loading ...
Loading ...
Select the channel to view and/or modify from the Channels list. This list indi
-
cates the channel number (with respect to this module) and the serial number of
the probe on each channel.
Click the Channel Settings button to display the appropriate Channel Settings
dialog. When this button is clicked, the settings for the selected channel are
read in from the thermometer readout. The Channel Settings dialog is described
in detail in Section 4.4.3.14, 1560 Black Stack PRT/Thermistor Channel Set
-
tings Dialog.
4.4.3.6.3 Calibration tab
Select the Calibration tab to view and/or modify module calibration informa
-
tion.
The settings on the Calibration tab may be passcode protected. If they are, a
message appears at the bottom of the Calibration tab and the Edit button is dis
-
played. The other controls on this tab are disabled. To edit the calibration pa
-
rameters, click the Edit button and enter the current passcode for the
thermometer readout. If the correct passcode is entered, the calibration parame
-
ter controls are enabled and the values may be changed. The controls remain
enabled until this dialog is closed or until the current changes are applied by
clicking the Apply button.
Caution: Do not change these values unless this module has been
recalibrated or unless these values do not match the values located on the
latest Report of Calibration for this module! Changing these values voids
the current calibration!
Select the date this module was last calibrated using the drop-down calendar or
by entering the date. This date should be located on the Report of Calibration
for this module.
Enter the values for the calibration constants in the boxes provided. The cali
-
bration constants required by the 2561 module are: 0 ADJ and 10 ADJ. The
107
4 Tools
Thermometer settings dialogs
Figure 79 2561 HPRT Module Settings - Calibration tab
1.888.610.7664 sales@GlobalTestSupply.com