Use the Digital Filter Enable drop-down list to select whether the digital filter
is enabled (On) or disabled (Off). If the setting is Off, the Average Count set
-
ting is not applicable and is disabled.
The Average Count set the number of Display Interval measurements that are
averaged by the digital filter. Possible values are between 2 and 100. If the Dig
-
ital Filter Enable is Off, this settings is not applicable and is disabled.
4.4.6.2 Display tab
Select the Display tab to view and/or modify display settings.
Select the units using the Units drop-down list. Readings can be displayed in
either °C, °F or K.
Note: This setting may be overridden depending on the selected probe
conversion type and channel settings for each channel. Please refer to the
thermometer readout's User's Guide for more information on setting the
system units.
Enter the resolution for temperature readings in the Temperature Resolution
box provided or click the up/down buttons. The temperature resolution can be
set to any value between 0 and 6 digits (places after decimal point).
User’s Guide
162
Figure 130 1594A/1595A Hart Super-Thermometer Display tab