Settings button is disabled. Please refer to Section 9, Conversion Settings
Dialogs, for a detailed description of the Conversion Settings dialogs.
Note: The conversion settings are not written to the thermometer readout
until the Apply or OK button is clicked on this dialog!
4.4.2.7.2 Calibration tab
Select the Calibration tab to view and/or modify channel calibration informa
-
tion.
Caution: Do not change these values unless this channel has been
recalibrated or unless these values do not match the values located on the
latest Report of Calibration for this channel! Changing these values voids
the current calibration!
Select the date this channel was last calibrated using the drop-down calendar or
by entering the date. This date should be located on the Report of Calibration
for this channel.
Select the date this channel is due for recalibration using the drop-down calen
-
dar or by entering the date.
Enter the values for the calibration constants in the boxes provided. The cali
-
bration constants required by a Thermocouple channel are: C0, C100 and CRJ.
The values of these constants should be located on the Report of Calibration for
this channel.
4.4.3 1560 Black Stack settings dialog
The 1560 Black Stack Settings dialog displays information about the thermom
-
eter readout including model number, serial number, identification information,
User’s Guide
96
Figure 66 1529 Chub-E4 thermocouple channel
settings - Probe tab