To use this software:
1. Click the Start button, then locate and click the icon created by the setup
process.
2. The software will start with the Log Session Settings dialog displayed. If
the Log Session Settings dialog is not displayed when the software starts,
select the Setup option in the Datalog menu to display the Log Session
Settings dialog.
Note: In order to make the Log Session Settings dialog appear automati
-
cally every time LogWare starts, check the “Show Log Session Settings di
-
alog on start-up” check box on the General tab of the Options dialog. The
Options dialog can be displayed by selecting the Options option in the
Tools menu. Please refer to Section 5.6.4, Options, for more details.
3. To start a log session, fill in the information on the Log Session Settings
dialog. The following information must be entered before a log session
can start:
Thermometer:
Select the model of the thermometer readout to use, enter the serial num-
ber and select the session temperature scale from the available choices.
The reading resolution may also be selected.
Select the COM port that the thermometer readout is connected to and
choose the baud rate to use.
Power up the thermometer readout and make sure it is connected to the
selected COM port and set to the selected baud rate. Click the Test Com
-
munication button to make sure LogWare can communicate with the ther
-
mometer readout.
Log session:
Select the interval between readings.
Optional - Enable and select the interval between scan sequences.
To set a delayed start date/time, click the Delayed Start button and choose
the appropriate date and time for the log session to start. Otherwise, the
log session will begin when the Start button is clicked.
Choose the appropriate end log session option and set the other end log
settings as required.
Channels:
Query the channels from the thermometer readout by clicking the Query
Channels button.
7
1 Introduction
Running LogWare
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com