4.4.3.2 Sampling tab
Select the Sampling tab to view and/or modify sampling settings.
Select the units to use when displaying readings in temperature using the Units
drop-down list. Temperature readings can be displayed in either °C, °F or K.
This setting applies to all channels that are configured to display readings in
temperature.
Select the time interval between subsequent measurements using the Sample in-
terval controls. The sample interval can be set to any value from 0 to 32,767
seconds (0 to 546 minutes or 0 to 9 hours).
Note: Setting the sample interval to 0 seconds or 1 second is equivalent to
setting it to 2 seconds because the 1560 Black Stack cannot sample at a
rate faster than once every 2 seconds. This is the default behavior of the
1560 Black Stack Thermometer readout.
Select the time interval between subsequent scans through a scan sequence us
-
ing the Scan sequence delay controls. The scan sequence delay can be set to
any value from 0 to 10,000 seconds (0 to 166 minutes or 0 to 2 hours).
Note: Setting the scan sequence delay to an interval of less than the set
-
ting of the sample interval multiplied by the number of channels in the
scan sequence effectively disables the scan sequence delay feature.
99
4 Tools
Thermometer settings dialogs
Figure 69 1560 Black Stack Settings – Sampling tab