9.9 Thermocouple Polynomial Settings dialog
The Thermocouple Polynomial Settings dialog allows the mV (25C), C0, C1,
C2,… C15 coefficients to be entered.
If the Thermocouple Polynomial coefficients were read in from the thermome
-
ter readout, the boxes are filled with the current coefficient values.
Enter the voltage of the thermocouple probe (in millivolts) at 25°C from the
Report of Calibration in the mV (25C) box. If this value is not applicable to the
thermometer readout, this box is disabled.
Enter the values for all thermocouple polynomial coefficients from the Report
of Calibration in the boxes provided. If any coefficients are not used, leave the
boxes blank or set the values to 0.0. The actual number of boxes displayed de
-
pends on the thermometer readout being used.
Select the Open button to display the Windows® Open dialog to open a previ
-
ously saved configuration file which contains Thermocouple Polynomial coeffi
-