User Manual for the Fluke 9935-S LogWare II Software - Page 127

For 9935-S.

PDF File Manual, 275 pages, Read Online | Download pdf file

9935-S photo
Loading ...
Loading ...
Loading ...
To close this dialog without writing the module settings to the thermometer
readout, click the Cancel button.
Click the Help button to open the appropriate help topic for this dialog.
4.4.3.9.1 General tab
Select the General tab to view and/or modify general module settings.
Select the excitation current for this module using the Excitation current
drop-down list. The 2564 module can use either a 2 µA or a 10 µA excitation
current, or Auto. This setting applies to all channels on this module.
Select the resistance range for this module using the Resistance range
drop-down list. The 2564 module allows the selection of either Auto, High or
Low. This setting applies to all channels on this module.
Select the front row wires option for this module using the Front row wires
drop-down list. The 2564 module allows the selection of either 4-wire or
3-wire. This setting applies only to the four channels on the front row of this
module.
Select the rear row wires option for this module using the Rear row wires
drop-down list. The 2564 module allows the selection of either 4-wire or
3-wire. This setting applies only to the four channels on the rear row of this
module.
User’s Guide
116
Figure 89 2564 THRS Module Settings - General tab
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com